Products

BT-500

Infant Incubator / Baby Incubator

Make:Bistos Koryea

BT-500 is an infant incubator, which can precisely control temperature and humidity.

BT-500 is an infant incubator, which can precisely control temperature and humidity with low noise
providing the best environment for the recovery of neonate. Also, CCD camera enables remote monitor-
ing of neonate, and in-built SpO2 Monitor and weighing scale provide total monitoring of neonate.
Based on ergonomic design, BT-500 will provide extreme comfort customers.

Specification

 

Physical characteristics

Standard

Dimension: 1,024(W)x690(D)x1,354(H)mm

Weight: Approx. 99.3 Kg

Full loaded

(including IV pole, IV plate, plate, external LCD monitor, shelf, weighing scale, CCD camera, lifting stand, O2 control module and check valve)

Dimension: 1,441(W)x797(D)x1,500(H)mm

Weight: Approx. 117.5 Kg

Safe working load: approx. 156.5 Kg

Body (Shell + Hood)

Dimension: 1,024(W)x604(D)x773(H)mm

Weight: Approx. 52.7 Kg

Fixed stand

Dimension: 1,009.5(W)x815(D)x645(H)mm

Weight: Approx. 36 Kg

Lifting stand

Dimension: 1,441(W)x797(D)x1,500(H)mm

Weight: Approx. 40.5 Kg

IV plate

Dimension: 479(W)x402(D)x756.4(H)mm

Weight: Approx. 5 Kg

Plate

Dimension: 402(W)x302(D)

Weight: Approx. 11 Kg

IV pole

Dimension: 300(W)x310(D)x718 ~ 1,068(H)mm

Weight: Approx. 2.5 Kg

External 7” color TFT LCD

Dimension: 364.7(W)x35(D)x839(H)mm

Weight: Approx. 3 Kg

Tilt angle: +30˚/ -50˚

Swivel angle: +180˚/ -140˚

Shelf

Dimension: 219(W)x334(D)x36(H)mm

Weight: Approx. 2.7 Kg

Weight limit: approx. 3 Kg

Basket

Dimension: 464.2(W)x490.5(D)x241.3(H)mm

Weight: Approx. 4.3 Kg

Weight limit: approx. 10 Kg

Basket partition

Dimension: 391(W)x205(D)x53(H)mm

Weight: Approx. 0.45 Kg

Weighing scale

Dimension: 810(W)x400(D)x46.6(H)mm

Weight: Approx. 5.8 Kg

Measurement range: 0 ~ 10 Kg, ±50g

CCD camera (NTSC or PAL)

Dimension: 69.1(W)x46(D)x38(H)mm

Weight: Approx. 40 g

Resolution: 510x492 pixel

Sensor module

Dimension: 157(W)x119.2(D)x64.5(H)mm

Weight: Approx. 270 g

Sensor module & CCD camera assembly

Dimension: 194(W)x119.2(D)x81(H)mm

Weight: Approx. 310 g

SpO2 probe

Wire length: 1,000 mm

Weight: approx. 30 g

SpO2 probe extension

Wire length: 3,080 mm

Weight: approx. 30 g

Mattress

Dimension: 727(W)x377(D)x27(H)mm

Weight: Approx. 30 g

Tilt angle: ±12˚

O2 control module

Dimension: 300(W)x170(D)x180(H)mm

Weight: Approx. 760 g

Check valve

Dimension: 330(W)x365(D)x65(H)mm

Weight: Approx. 460 g

 

 

 

The following materials are used

Mattress

Polyurethane

 

 

 

 

Functional Characteristics

Air temperature: control and measurement of air temperature in the hood

Control range

Normal mode: 23.0 ℃ ~ 37.0

Override mode: 37.1 ℃ ~ 39.0 

Measurement range

20.0 ℃ ~ 45.0 

Accuracy

± 0.5 

Skin temperature: control and measurement of skin temperature of the infant

Control range

Normal mode: 35.0 ℃ ~ 37.5

Override mode: 37.6 ℃ ~ 39.0 

Measurement range

25.0 ℃ ~ 45.0 

Accuracy

± 0.5 

Accuracy of skin temperature sensor

± 0.3 

Humidity: control and measurement of relative humidity in the hood

Control system

Steam humidifier

Control range

40 %RH ~ 95 %RH

Measurement range

15 %RH ~ 99 %RH

Accuracy

± 5 %p

Autoclavable water box

 

O2: measurement of O2 in the hood

Measurement range

18 % ~ 100 %

Accuracy

± 5 %p

O2: servo control of O2 in the hood

Control range

21 % ~ 65 %

Accuracy

± 5 %p

Max inlet pressure

120 psi

Hose diameter

8mm, 9.5mm, 11mm, 13mm (Use only oxygen hose)

Weighing scale: measurement of weight of the infant

Measurement range

0.000 Kg ~10.000 Kg

Accuracy

± 50 g

Pulse rate(PR): measurement of pulse rate of the infant

Measurement range

30 bpm ~ 240 bpm

Error tolerance (Without movement noise)

± 3 digit

Resolution

1 bpm

SpO2: measurement of SpO2 of the infant

Measurement range

1 % ~ 100 %

Accuracy (Without movement noise)

70 % ~ 100 %, ± 3 digit

0 % ~ 69 %, unspecified

Resolution

1 %

Power (AC)

Internal battery

Input

AC 100/240, 50/60Hz

Li-ion rechargeable battery (3.7V, 2600mAh)

Consumption

1200VA

5 hours to full charge when main power switch is on.

Alarm sounds for 30 minutes while no power

Operation environment

Storage environment

Temperature

20 ~ 30°C (68 ~ 86°F)

Temperature

–20 ~ 60°C (4 ~ 140°F)

Humidity

0 ~ 95% non-condensing

Humidity

0 ~ 95% non-condensing

Air pressure

70~106 kPa

Air pressure

70~106 kPa

Altitude

0 - 2,000m(0 – 6,561.68 ft)

Altitude

0 - 2,000m(0 – 6,561.68 ft)

Other environment

 

Maximum carbon dioxide(CO2) concentration

<0.5% at a point 15cm above from the center of the mattress

Air velocity over the mattress

<0.3 m/s

Warm-up time

Normal mode: 45 minutes, <55 dBA

Silent mode:  60 minutes, <45 dBa

Fast mode:   30 minutes, >66 dBA

Sound pressure level in hood

Under 45 dBA in environmental sound pressure level 40 dBA

Standard

 

Complies with IEC60601-1, IEC60601-1-2, IEC60601-2-19 and ISO 80601-2-61

Class I equipment & Internal powered equipment

Continuous operation

Type BF applied part

Skin temperature probe: IPX1

SpO2 probe: IPX2

Foot switch: IPX6

Alarm signal sound pressure

 

Priority

High (All)

Measured sound pressure level

73 dBA

A-weighted background level

48 dBA

 