Products

BM7

Patient Monitors

Make:Bionet Korea

Features

Parameter versatility from recovery to post-op
  • BM7 offers wide application coverage from recovery ward up to ICU, pre-op, post-op, & quick surgery application with Bionet’s engineering BMGA Multi-gas analyzer as well as main and side stream EtCO2 capnography.
User friendly display design & Interface
  • 12.1″ LCD touch screen that displays clear and enlarged parameters, allowing care-givers to get access to operational measurement in tip of their fingers.
Pop up menu
  • All Bionet patient monitors offer Pop-up trend for ECG events with ST segment analysis for easy recognition.
INOP alarm
  • Adds operational values with informing users to acknowledge immediate physiological & technical alarms of IEC60601-1-8 requirement.
Excellent network connectivity
  • Features care-givers’ workflow demand from patient admission to discharge with barcode readers and ID by direct HL7 interface. Its central connectivity allows up to 64 monitors (16 wireless monitors) that permits users to manage patient data with remote NIBP triggering, alarm settings, and exporting to HER
Parameters Product Specifications Value
Hardware specification Dimension (H x W x D) 245 x 322 x 225 mm
  Weight Approx. 4.5kg
  Power AC 100-240V (50/60Hz), Adapter 18 V, 2.8 A
  Display 12.1"color TFT LCD
  Resolution 1024 x 768 pixels
  Parameters ECG, Heart Rate, Respiration Rate, SpO2, Pulse Rate, Systolic BP, Diastolic BP, Mean BP, 2 x Temperature, 2 x IBP, EtCO2, FiCO2, Airway Respiration Rate, Anesthetic Gas Concentration in inhalation and exhalation
  Traces 6 waveforms : 2*ECG, SpO2, RR or EtCO2 or MGAS, 2*IBP
Sweep speed : 6.25, 12.5, 25, 50 mm/sec
  Indicators Categorized alarms (3 priority levels)
Visual alarm lamp handle
QRS beep & SpO2 pulse beep
Percent(%) SpO2 pitch tone
External power LED, Battery status
Touch screen, rotary knob
  Interfaces DC input connector: 18VDC, 2.8A
LAN digital output for transferring data
Nurse call system connection:
• 0.3A at 125VAC
• 1A at 24VDC
DC output: 5VDC, 1A (Max)
External monitor connection: HDMI
USB barcode reader
USB data storage
  Battery Rechargeable Li-ion battery
2 hours for continuous operation
(Option: using additional battery, total of 4 hours)
  Thermal Printer (optional) Speed : 25, 50mm/sec
Paper width : 58mm
  Data Storage 168 hours trends data, 20 cases of 10 sec alarm waveform
  Language English, French, Spanish, Italian, Germany, Chinese, Russian, Czech, Bulgarian, Portuguese, Romanian, Hungarian, Turkish, Polish, Korean
Environment Temperature Operation: 5 ~ +40 °C (41 ~ 104 °F)
Storage: –20 ~ +60 °C (–4 ~ +140 °F)
  Humidity Operation: 30% ~ 85%,
Storage: 10% ~ 95% (Package )
  Altitude Operation: 525 ~ 795 mmHg (70 ~ 106 kPa)
Storage: 375 ~ 795 mmHg (50 ~ 106 kPa)
ECG Lead type 3-lead, 5-lead
  Lead Selection 3-lead : I, II, III
5-lead : I, II, III, aVR, aVL, aVF, V
  ECG waveforms 3-lead : 1 channel
5-lead : 2/7 channels
  Heart Rate Range Adult : 30 – 300 bpm
Neonate/Pediatric : 30 – 350 bpm
  Heart Rate Accuracy ±1bpm or ±1%, whichever is greater
  Sweep speed 6.25, 12.5, 25, 50 mm/sec
  Filter Diagnosis: 0.05Hz ~ 150Hz
Monitoring: 0.5 ~ 40 Hz
Moderate: 0.5 ~ 25 Hz
Maximum: 5 ~ 25 Hz
  S-T segment detection range -2.0 to 2.0 mV
  Arrhythmia analysis ASYSTOLE,VTACH,VFIB,BIGEMINY,ACCVENT, COUPLET,IRREGULAR, PAUSE,PVC,RONT,TRIGEMINY,VBRADY, SHORTRUN
  Pacemaker Detection Mode Indicator on waveform display (user selectable)
  Protection Against electrosurgical interference and defibrillation
Respiration Performance Method Thoracic impedance
  Channel selection RA-LA or RA-LL
  Measurement range 5 – 120 breaths per minute
  Accuracy ±1 breath per minute
  Apnea alarm Yes
SpO2 Performance Saturation range 0 to 100%
  Saturation accuracy 70 to 100% ±2 digits
0 to 69% unspecified
  Pulse rate range 30 to 254 bpm
  Pulse rate accuracy ±2 bpm
NIBP Performance Method Oscillometric
  Operation Mode Manual/Automatic/Continuous
NIBP Performance Measurement range Systolic : 40 to 260 mmHg
MAP : 26 to 220 mmHg
Diastolic : 20 to 200 mmHg
  Accuracy mean error: less than ±5 mmHg
standard deviation : less than 8 mmHg
Meets accuracy requirements of ANSI/AAMI SP10:1992 and 2002 (*SunTech)
Temperature Performance Measurement range 0 to 50℃ (32 to 122℉)
  Accuracy 25 to 50℃: ±0.1℃
0 to 24℃: ±0.2℃
  Compatibility YSI Series 400 temperature probes
IBP Performance (Optional) Channels 2
  Measurement range -50 to 300mmHg
  Accuracy Less than 100mmHg : ±1mmHg
Equal or greater than 100mmHg : ±1% of reading
  Pulse rate measurement range 0 to 300 BPM
  Zero balancing Range : ±200mmHg
Accuracy : ±1mmHg
Drift : ±1mmHg over 24 hours
  Transducer sensitivity 5µV/V/mmHg
Sidestream CO2 (Optional) Measurement range 0 to 150 mmHg, 0 to 19%
  Accuracy 0-40mmHg ±2 mmHg,
41-70mmHg ±5% of reading
71-100mmHg ±8% of reading,
101-150mmHg ±10% of reading
  Respiration rate 2 to 150 breaths per minute
  Respiration accuracy ±1 breath per minute
  Measurement range 0 to 150 mmHg, 0 to 19%
  Accuracy 0-40mmHg ±2 mmHg,
41-70mmHg ±5% of reading
71-100mmHg ±8% of reading,
101-150mmHg ±10% of reading
  Respiration rate 0 to 150 breaths per minute
  Respiration accuracy ±1 breath per minute
Anesthetic GAS/O2 - Phasein (Optional) Method Infra-red absorption characteristic
  Gas CO2, O2, N2O, Des, Iso, Enf, Hal, Sev
  Warm-up time Main stream (IRMA AX+):
• Iso accuracy mode: 45s
• Full accuracy mode : 60s
Side stream (ISA OR+/ AX+) : <20s
  Sample flow rate (for ISA OR+/AX+) 50 ± 10 ml/min
  Measuring Range CO2 : 0 ~ 15 %
N2O: 0 ~ 100%
Hal/Iso/Enf: 0 ~ 8%
Sev: 0~10%
Des: 0 ~ 22%
O2: 0 ~ 100%(ISA OR+/AX+)
  Respiratory Rate 0 ~ 150bpm ± 1bpm

Copyright © 2001-2020 MKJ Associates. All Rights Reserved.