Your position:Home> Products>>

Products

BSC-1000IIA2

Class II A2 Biological Safety Cabinet

Make:DRAWELL

Class II A2 Biological Safety Cabinet Model BSC-1000IIA2
 
Working principle
The bio-safety cabinet is a box-shaped air purification negative pressure safety device that prevents aerosols from containing dangerous or unknown biological particles during handling.
 
Its working principle is mainly to pump the air inside the cabinet to keep the negative pressure state inside the cabinet to protect the operator through the vertical air flow. The outside air is filtered through the HEPA filter and then enters the safety cabinet to avoid contamination of the sample. The air in the cabinet is also filtered through the HEPA filter and then vented to the atmosphere to protect the environment.
 
Characteristics
1.10° tilt design, more ergonomic
2. Vertical flow negative pressure machine type, 30% of air is filtered and recycled, 70% air is filtered and then discharged into the room or connected to the exhaust system.
3. Lighting and sterilization system safety interlock
4.HEPA high-efficiency filter can filter more than 99.99% of 0.3um dust particles
5. Digital display LCD control interface, fast, medium. Slow three speed
6. The working area is SUS304 brushed stainless steel, strong, durable, easy to clean, anti-corrosion
9. Standard configuration diameter 160mm, 1 meter long exhaust duct and elbow
10. Work area is equipped with a five-hole socket
 
Technical Parameters

Model BSC-1000IIA2
Clean level Grade 100(209 E U.S. Federal)
Number of bacteria ≤0.5per utensil.hour(90mm utensil)
Average wind speed Inside the door 0.38±0.025 m/s
Middle 0.26±0.025 m/s
Inside 0.27±0.025 m/s
Front suction wind speed 0.55m±0.025m/s (70% air exhaust)
Noise ≤62dB(A)
Working area dimension(W*D*H) 1000*650*620mm
Overall dimension(W*D*H) 1195*720*1950mm
HEPA filter specifications and quantity 955*554*50mm*1pc
Fluorescent/UV lamp specifications and quantity 20W*1PC/20W*1PC

Copyright © 2001-2020 MKJ Associates. All Rights Reserved.