Products
-
- Operating Theatre Equipment
- OB/GYN Equipment
- Diagnostic Equipment
- Laboratory Equipment
- Biological Microscope
- Spectrophotometer
- High Speed Centrifuge
- Low Speed Centrifuge
- High Speed Freeze Centrifuge
- Low Speed Freeze Centrifuge
- Centrifuge
- Thermostatic Bath
- Laboratory Shaker
- Laboratory Mixer
- Water Distiller
- Stirrer
- Laboratory Incubator
- Dry Oven
- PH Meter
- Biological Safety Cabinet
- Laminar Flow Cabinet
- Water Purifier
- Ultrasonic Cleaner
- Pipette
- Microtome
- Medical Freezer
- Blood Bank Refrigerator
- Medical Pharmacy Refrigerator
- Balance
- Medical Destroyer
- Blood Bag Tube Sealer
- Furnace
- Hospital Furniture
- First-Aid Products
- Ophthalmic Equipment
- ENT Equipment
- Home Care Equipment
Main Features:
* Horizontal slit optical design, with excellent luminous detection efficiency and high signal to noise ratio. Water Raman peak SNR better than 350: 1 (P-P) or 1500: 1 (RMS). The minimum sample volume is 0.5mL when using a standard 10mm square sample cell.
* Wavelength scanning speed up to 60000 nm / min.
* Bandwidth 1/2/5/10/20nm adjustable.
* Fluorescence Quantum Yield accessory available.
* The software offers fluorescence 3D scanning, equal-wavelength difference synchronous scanning, equal wave number difference (constant energy difference) synchronous scanning functions. (synchronous scanning function require pro version software)
* Multi accessories available.
* Using Hamamatsu’s high quality Xenon light source and photoelectric multiplier tube detectors and instruments to provide sufficient light intensity signal and the detection sensitivity.
* Built-in frequency filter. Built-in optical gate, designed for unstable sample.
* Package size:785*665*530mm
* G.W:40kg
Model |
Specification of F98 Series Fluorescence Spectrophotometer |
Excitation Source |
150W xenon lamp(Hamamatsu) |
Excitation Wavelength |
200nm~900nm |
Emission Wavelength |
200nm~900nm |
Excitation Slit |
1nm/2nm/5nm/10nm/20nm |
Emission Slit |
1nm/2nm/5nm/10nm/20nm |
Wavelength Accuracy |
±0.4nm |
Wavelength Repeatability |
≤0.2nm |
Signal-to-Noise Ratio |
S/N≥350(P-P) S/N≥1500(RMS) S/N≥15000(RMS BG) |
Limit |
≤5×10-11 g/ml(Quinine Sulfate Solution) |
Linearity |
γ≥0.995 |
Peak Repeatability |
≤1.5% |
Wavelength Scan Speed |
Multi-speed Level, Maximum at 60000nm/min |
Minimum sample size |
0.5mL (Using a standard 10mm square sample cell) |
Octave filter |
Build in Octave filter |
Photometric QuantityRange |
-9999 ~ 9999 |
Response Time |
0.02/0.1/0.5/1/2/4/8 s,Auto adjust |
Data Transportation |
USB2.0 |
Power |
200W |
Power Source |
AC 220V/50Hz;110V/60Hz |
Instrument Dimension |
610×460×365(mm) |
Weight |
Net Weight:21kg Gross Weight:26kg |