BT-550

Infant Warmer

Make:Bistos Koryea

Infant warmer BT-550 is intended to emit controlled, evenly distributed overhead heat to the body of premature infant and other new-borns who cannot effectively regulate their body temperature. This device can be used before any treatment or operation of infant.

The BT-550 is intended to emit controlled, evenly distributed overhead heat to the body of premature infant and other newborns that cannot effectively regulate their body temperature. This device can be used before any treatment or operation of infant. The device has two operating functions; one is baby mode, and the other one is manual mode. In the baby mode, the temperature is controlled by skin temperature of infant. The skin temperature of infant compared to the temperature setting. The set value is lower than the measured value the heater of device is operate to increase the value. In the manual mode, the device is operates for predefined time.
The device has the option for weighing scale and SpO2 measuring functions. Weight and SpO2 of infant can be measured using the options. In the SpO2 measurement, the sensor is place at the end of a finger. By measuring the intensity of reflected light, depending on the concentration of dissolved oxygen in the blood determines the oxygen concentration.

Specification

 

Functional Characteristics

Temperature

Infrared Heater

Operation Mode

(Each of tolerance range is ± 20%)

-Prewarm (100% @28/)

-Baby mode (100% @ 28 /)

-Manual (100% @ 8 /)

Manual Control Range

0 ~ 100%, 20levels

Skin Display Range

26 ~ 42°C ±0.3°C

Output Power

600W

Skin Control Range

34 ~ 38°C (in Baby Mode)

 

 

Display

APGAR

Type

7" TFT Color LCD

Setting Range

0sec ~ 59min 59sec

Alarm
(Visual & Sound)

6 Events Alarms

Indicating Beep

1, 5, 10min

Examination Lamp

Function

LED

40W(10W x 4ea)

Microprocessor Controlled 

Brightness
Control

3levels

Rotation of Head

90° (Left, Light both Directions)

Illumination

> 7,000lx

Tripod Water Level

 

Power

PC Interface

Input

AC 100/240, 50/60Hz

RS232C

 

Consumption

750VA

 

 

Standard Configuration (Fixed height stand)

Skin Temperature Sensor

2ea

Mattress

Quick guide

1ea

1ea

Power Cord

1ea

Operation manual

1ea

Options

Weight limit or Angle

Options

Weight limit or Angle

Lifting Stand

 

IV Pole

Approx. 5 kg

IV Plate

Approx. 11 kg

Weighing Scale

0 ~ 10.0 kg, ±50g

Basket(Drawer)

Approx. 10 kg

Masimo SpO2

 

Basket Partition

 

Extension for SpO2

 

Tilting Bed

±15°

   

Physical Characteristics

Dimension

Weight

Standard 

1,027(W)x690(D)x1,890(H)mm

Standard

83 Kg

Fixed  stand

Lifting stand

725(H) mm

615 ~ 815(H)mm

Full option

98 kg

Mattress

495(W) x 810(D) x 27(H)mm

with Safe Working Load

Approx. 124kg 

Packing

1,150(W)X750(D)x1,550(H)mm

Packing

126Kg(Full Option)

Operation environment

Storage environment

Temperature

18 ~ 30°C (64.4  ~ 86°F)

Temperature

–10 ~ 60°C (14 ~ 140°F)

Humidity

0 ~ 95% non-condensing

Humidity

0 ~ 95% non-condensing

Air pressure

70~106 kPa

Air pressure

70~106 kPa

Standard

 

IEC60601-1, IEC60601-1-2

IEC60601-2-21, ISO 80601-2-61

 

   
     
               

 

 

 

 

Masimo Oximeter

Range

Saturation (% SpO2)

Pulse Rate (bpm)

Perfusion

1%  - 100%

25 - 240

0.02% - 20%

Accuracy

Saturation (% SpO2) – During no motion conditions

Neonate

70% - 100%, ± 3 digits

0% - 69%, unspecified

Saturation (% SpO2) – During motion conditions

Neonate

70% - 100%, ± 3 digits

0% - 69%, unspecified

Pulse Rate (bpm) – During no motion conditions

25 to 240, ± 3 digits

Pulse Rate (bpm) – During motion conditions

25 to 240, ± 5 digits

Resolution

Saturation (% SpO2)

Pulse Rate (bpm)

1%

1

Low Perfusion Performance

> 0.02% Pulse Amplitude

 and % Transmission > 5%

Saturation (%SpO2) ± 2 digits

Pulse Rate ± 3 digits

Interfering Substances

Carboxyhemoglobin may erroneously increase readings. The level of increase in approximately equal to the amount of carboxyhemoglobin present. Dyes or any substances containing dyes that change usual arterial pigmentation may cause erroneous readings.

Power

Voltage Input Range

Max. AC power consumption

Rechargeable battery

90 – 240 VAC, 47 – 63 Hz

20 VA

Up to 7 hours battery life

Fuses

0.75A, Time Delay, 250 V

 

Isolation

Chassis Leakage current

Ground resistance

Less than 100 mp

Less than 1.0 Ω

Environment

Operating temperature

Storage temperature

Relative humidity

Operating altitude

41 to +104(5 to + 40)

-40 to +158(-40 to + 70)

5% to 95% non-condensing

500 mbar to 1060 mbar pressure

-1,000ft to 18,000ft(-304m to 5,486m)

Circuitry

Microprocessor controlled

Automatic self-test of oximeter when power on

Automatic setting of default parameters

Automatic alarm messages

Trend data output of SpO2, pulse rate

Auto indicators

Adjustable volume audible pulse: OFF and 33% to 100% in 4 steps

Adjustable volume audible alarm tone: levels and 33% to 300% in 4 steps

Alarm silence (120 seconds); all mute (continuous silence)

Smart tone ON/OFF: excessive motion and low perfusion conditions pulse tones

Pulse rate out-of-limits alarm

SpO2 level out-of-limits alarm

Sensor conditions alarms

System failure and battery low alarms

Physical characteristics

Dimensions

 

Weight

8.2” x 6.0” x 3.0”

(20.8cm x 15.2cm x 7.6cm)

2.1 lbs, 32 oz (0.928 kg)

Modes

Averaging mode

Sensitivity

2,4,8,10,12,14 and 16 seconds

Normal, APOD, and MAX

 

 